Dataset for multiple publishable figures in the paper entitled "Leakage Current Pathways in Josephson Arbitrary Waveform Synthesizer Standards" to be submitted to CPEM 2024

The voltage errors associated with leakage currents in Josephson arbitrary waveform synthesizer (JAWS) systems are significant contributors to the overall system accuracy. This paper describes discrepancies in output voltage between two different circuit halves on a single JAWS chip and shows that this discrepancy is dominated by ac leakage currents through the stray capacitance in the compensation leads.
Organization: National Institute of Standards and Technology
Last updated: 2024-03-12T23:56:10.076900
Tags: josephson-junctions, leakage-currents, measurement-uncertainty, standards, superconducting-integrated-circuits, voltage-measurement

Tables

Fig3

@usgov.national_institute_of_standard_leakage_current_pathway_475e2029.fig3
  • 5.83 KB
  • 7 rows
  • 6 columns
Loading...

CREATE TABLE fig3 (
  "ac_ac_difference_between_jaws_a_and_jaws_b_measurement_849d6036" VARCHAR,
  "unnamed_1" VARCHAR,
  "unnamed_2" VARCHAR,
  "unnamed_3" VARCHAR,
  "unnamed_4" VARCHAR,
  "unnamed_5" VARCHAR
);

Table 1

@usgov.national_institute_of_standard_leakage_current_pathway_475e2029.table_1
  • 5.44 KB
  • 7 rows
  • 6 columns
Loading...

CREATE TABLE table_1 (
  "the_jaws_chip_had_two_halves_a_and_b_each_containing_f_81055477" VARCHAR,
  "unnamed_1" VARCHAR,
  "unnamed_2" VARCHAR,
  "unnamed_3" VARCHAR,
  "unnamed_4" VARCHAR,
  "unnamed_5" VARCHAR
);

Share link

Anyone who has the link will be able to view this.